Video instruction
Text instruction

Preparation
If the test was stored in the refrigerator, allow all materials (including the test cassette and buffer) to reach room temperature (15–25°C / 59–77°F) before testing.
Prepare a clean and flat working surface.
Have all materials ready before starting the test.

Place Cassette
Take out the test card from the foil pouch and place it horizontally.

Sample Collection
Use the provided lancet to gently prick the dog's ear, paw or inside of the mouth.

Collect Sample
Use Pipette to collect whole blood (fresh or stored at 2–8°C for use within 3 days).

Mix the Sample
Drop 1 drop into a vial of assay buffer and mix well.

Apply Sample
Add 4 drops of the diluted sample into each sample hole of the test.

Wait and Read Results
Wait 5–10 minutes before reading the result. Do not interpret results after 10 minutes, as they are considered invalid.

Positive Results
Both the C band and the T band are visible, regardless of whether the T band is strong or faint.

Negative Result
Only a clear C band is visible.

Invalid Result
No colored C band appears, regardless of whether the T band is visible.
A retest is recommended.